氧化锆角度基台及种植体周骨壁应力分布的有限元分析

杨炎忠1, 田小华1, 周延民2   

  1. 1.杭州口腔医院,浙江杭州 313000; 2.吉林大学附属口腔医院,吉林长春 130041
  • 收稿日期:2014-07-02 出版日期:2015-08-20 发布日期:2015-09-10
  • 通讯作者: 周延民,Tel:0431-88796025,E-mail:zhouym@jlu.edu.cn E-mail:zhouym@jlu.edu.cn
  • 作者简介:杨炎忠(1979-),男,硕士,E-mail:15167134818@163.com

Effect of zirconia abutment angulation on stress distribution in the abutment and the bone around implant: a finite element study

YANG Yan-zhong1, TIAN Xiao-hua1, ZHOU Yan-min2   

  1. 1.Hangzhou Stomatological Hospital. Hangzhou 313000, Zhejiang Province; 2.Stomatological College of Jilin University. Changchun 130041, Jilin Province, China
  • Received:2014-07-02 Online:2015-08-20 Published:2015-09-10

摘要: 目的研究氧化锆角度基台及种植体周骨壁应力的分布情况,为临床应用提供参考。方法采用三维有限元分析方法,对氧化锆直基台、15°、20°基台修复时的种植体基台、固定螺丝及种植体周骨壁应力的分布进行分析,并与常规钛基台进行比较。结果角度基台模型的基台、螺丝、种植体周骨壁的等效应力大于直基台;20°基台模型的基台、螺丝、种植体周骨壁等效应力远远大于15°基台模型;氧化锆基台模型与钛基台模型相比,2组无显著差异。结论基台角度对应力分布有影响,并随角度的增大而增大,提示临床上应注意种植体植入方向,尽量使用直基台或小角度基台,以减少种植体颈部骨吸收及修复并发症的发生;基台材料改变对基台、螺丝、种植体周骨壁的应力分布无显著影响,从生物力学考虑,临床上可以放心选用氧化锆角度基台。

关键词: 氧化锆, 基台, 角度, 种植体

Abstract: PURPOSE: To investigate the effect of three different zirconia angular abutments on the stress distribution in bone and abutment using three-dimensional finite element analysis, and provide instruction for clinical application. METHODS: Finite element analysis (FEA) was applied to analyze the stress distribution of three different zirconia/titanium angular abutments and bone around implant. RESULTS: The maximum Von Minses stress that existed in abutment, bolt and bone of the angular abutment model was significantly higher than that existed in the straight abutment model. The maximum Von Minses stress that existed in abutment, bolt and bone of the 20° angular abutment model was significantly higher than that existed in 15° angular abutment model. There was no significant difference between zirconia abutment model and titanium abutment model. CONCLUSIONS: The abutment angulation has a significant influence on the stress distribution in the abutment, bolt and bone, and exacerbates as the angulation increases, which suggest that we should take more attention to the implant orientation and use straight abutment or little angular abutment. The zirconia abutment can be used safely, and there is no noticeable difference between zirconia abutment and titanium abutment on stress distribution.

Key words: Zirconia, Abutment, Angulation, Implant Shanghai J Stomatol, 2015, 24(4):447-450.