高压蒸汽灭菌对3种机用镍钛器械形貌及抗力的影响

李向芬,郑平,徐丽,苏勤   

  1. 口腔疾病研究国家重点实验室四川大学,四川成都610041
  • 收稿日期:2015-03-25 出版日期:2015-12-25 发布日期:2016-01-04
  • 通讯作者: 苏勤,Tel:028-85501439,E-mail:susan_sq@hotmail.com E-mail:lixf9fighting@sina.com
  • 作者简介:李向芬1991-,女,硕士

The influence of autoclave sterilization on surface characteristics and cyclic fatigue resistance of 3 nickel-titanium rotary instruments

LI Xiang-fen, ZHENG Ping, XU Li, SU Qin.   

  1. State Key Laboratory of Oral Diseases, Sichuan University. Chengdu 610041, Sichuan Province, China
  • Received:2015-03-25 Online:2015-12-25 Published:2016-01-04

摘要: 目的 比较高压蒸汽灭菌对3种机用镍钛器械(K3、Mtwo、ProTaper)形貌及循环疲劳抗力的影响。方法 分别选取3种不同品牌的机用镍钛器械K3、Mtwo、ProTaper各36根。首先取各品牌器械24根,分为4组(n=6),分别进行循环高压蒸汽灭菌0、1、5、10次后,在人工根管内自由旋转至器械折断,记录折断前的旋转时间和折断尖端长度,采用SPSS13.0 软件包对数据进行单因素方差分析,比较其循环疲劳抗力的差别。再将剩余的器械各12根分为2组(n=6),第1组不予处理,第2组循环高压蒸汽灭菌10次,扫描电镜观察各器械尖端外表面形貌及横截面形貌的改变。结果 未灭菌组中,循环疲劳抗力K3大于Mtwo,并大于ProTaper(P<0.05);K3组5、10次组较未灭菌组均表现出抗力增强趋势,仅10次组与未灭菌组间差异有显著性(P<0.05);ProTaper组所有灭菌组较未灭菌组均显著增强(P<0.05),其中,5次组抗力最大;Mtwo组中,随着灭菌次数的增加,循环疲劳抗力逐渐增强(P<0.05)。扫描电镜观察发现,镍钛器械外表面和内部的不规则缺陷在多次灭菌后均显著增加。结论 不同品牌器械之间循环疲劳抗力不同,高压蒸汽灭菌在一定程度上可增加镍钛器械的循环疲劳抗力,但却使器械表面的粗糙度增加,横截面微观缺陷增多。

关键词: 镍钛器械, 灭菌, 循环疲劳, 器械分离

Abstract: PURPOSE To investigate the effects of autoclave sterilization on surface characteristics and cyclic fatigue resistance of 3 types of nickel-titanium rotary instruments (K3, Mtwo, ProTaper). METHODS Three brands of NiTi rotary endodontic instruments of the same size (tip diameter 0.25 mm and constant 0.06 taper) were selected: K3, Mtwo and Protaper (F2). 24 instruments for each brand were used to evaluate the effects of autoclave sterilization on inner character in the as-received condition and after subjection to 0, 1, 5, and 10 sterilization cycles (6 for each group). Time to fracture (TtF) from the start of the test to the moment of file breakage and the length of the fractured fragment were recorded. Means and standard deviations of TtF and fragment length were calculated. The data was analyzed with SPSS13.0 software package. Another 12 NiTi rotary instruments for each brand were used, 6 subjected to 10 autoclave sterilization cycles and the other as control. Scanning electron microscope was used to observe the changes in surface topography and inner character. RESULTS:For cyclic fatigue resistance, when sterilization was not performed, K3 showed the highest value of TtF means and ProTaper the lowest. The differences between each brand were statistically significant (P<0.05).When disinfection was performed, K3 brand showed greater fatigue resistance in comparison with the control when autoclave sterilization cycled 5 times and 10 times. The difference between 10 cycles of sterilization and the control was statistically significant (P<0.05); ProTaper brand showed significantly greater fatigue resistance in all the disinfected groups compared with the control (P<0.05) and 5 cycles of sterilization led to the greatest increment; The fatigue resistance of Mtwo brands increased with sterilization cycles and the difference between 5/10 cycles and the control were statistically significant (P<0.05).For surface characteristics, under scanning electron microscope, surface and inner imperfections in all instruments were intensified greatly after 10 cycles of sterilization.CONCLUSIONS Cycle fatigue resistance is different among instruments of different brands. Autoclave sterilization may increase fatigue resistance of the 3 brands. Autoclave sterilization may increase the surface roughness and inner defects in cross section.

Key words: Nickel-titanium rotary instruments, Autoclave sterilization, Fatigue resistance, Instrument fracture